Abstract

In cancer therapy using an ion beam, a scanning irradiation method has been developed in order to realize 3-D conformal irradiation. From studies of the scanning method, it has been required for a medical synchrotron to be capable of controlling the beam intensity during the duration of extraction. For this purpose, a beam-intensity control technique with RF-knockout slow-extraction has been studied and developed at the HIMAC synchrotron. Based on a simple one-dimensional model of extraction, the function of amplitude modulation (AM) for a transverse RF-field was analytically calculated to control the beam current. By cooperating with the feedback system, finally, we could control the beam intensity within a range of one order during a single flattop. We report on a simulation and the experiment result of intensity control for scanning irradiation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.