Abstract

Otitis media with effusion is a worldwide ear disease. The current treatment is to surgically insert a ventilation grommet into the tympanic membrane. A robotic device allowing automatic grommet insertion has been designed in a previous study; however, the part of the membrane where the malleus bone is attached to the inner surface is to be avoided during the insertion process. This paper proposes a synergy of optical flow technique and a gradient vector flow active contours algorithm to achieve an online tracking of the malleus under endoscopic vision, to guide the working channel to move efficiently during the surgery. The proposed method shows a more stable and accurate tracking performance than the current tracking methods in preclinical tests. With satisfactory tracking results, vision guidance of a suitable insertion spot can be provided to the device to perform the surgery in an automatic way.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.