Abstract
The inconvenience and risk associated with the regular use of invasive blood glucose measurements has led to tremendous research in this area. This paper proposes the design of a non-invasive blood glucose estimation system using novel Mel frequency cepstral coefficients features of wristband photoplethysmogram signal and physiological parameters. A dataset from 217 participants of a hospital in Cuenca Ecuador is used to validate the proposed model. The support vector regression (SVR) and extreme gradient boost regression (XGBR) techniques are used to estimate blood glucose levels (BGL). The XGBR technique achieves the least value for the standard error of prediction (SEP), 9.78 mg/dL. Further, 5 features are selected from the feature set based on the feature importance in XGBR. The XGBR model with the reduced feature set results in further reduction of SEP value (5.53 mg/dL) with a correlation coefficient of 0.99. Standard Clarke error grid analysis and Bland-Altman analysis shows that the predicted glucose values are in the clinically acceptable region. The results of the proposed model demonstrate the potential of wearable BGL monitoring technology.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.