Abstract

A rapid, reliable and precise integrated solid-phase extraction (SPE) and reversed-phase liquid chromatography method was developed and validated to determine bupivacaine in human serum using single protein-coated analytical columns. The protein-coated columns were packed with four different sorbents: TSK-ODS, LiChrosorb RP-8, LiChrosorb RP-2 and μ-Bondapak CN-bonded silica. The method involved direct injection of serum sample onto the columns for trapping of the analyte, clean-up from weakly retained serum endogenous components, as well as the final separation. The protein-coated columns operated in two different chromatographic modes. Serum proteins were extracted and cleaned up by SPE, whereas the final separation of bupivacaine was based on reversed-phase chromatography. The protein-coated TSK-ODS column resulted in more accurate peak integration and more reproducible results. A linear relationship between the concentrations of drug and peak areas was confirmed in the range of 100-2000ng/mL. Detection and quantification limits were 24.85 and 85.36ng/mL, respectively. The average recovery for bupivacaine ranged from 96.48% to 98.81%. The present methodology was successfully applied, with a high degree of confidence, to analyze clinical samples obtained from patient receiving 0.5% bupivacaine therapy.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call