Abstract
This work presents development of an integrated ultrasound (US)–cone-beam CT (CBCT) system for image-guided needle interventions, combining a low-cost ultrasound system (Interson VC 7.5MHz, Pleasanton, CA) with a mobile C-arm for fluoroscopy and CBCT via use of a surgical tracker. Imaging performance of the ultrasound system was characterized in terms of depth-dependent contrast-to-noise ratio (CNR) and spatial resolution. US–CBCT system was evaluated in phantom studies simulating three needle-based procedures: drug delivery, tumor ablation, and lumbar puncture. Low-cost ultrasound provided flexibility but exhibited modest CNR and spatial resolution that is likely limited to fairly superficial applications within a ∼10cm depth of view. Needle tip localization demonstrated target registration error 2.1–3.0mm using fiducial-based registration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.