Abstract
A package of two planar gradient coils combined with an RF coil was microfabricated for low-field MRI measurements of velocity flow profiles. The package generates orthogonal gradients in the radial and velocity encoding directions for standard flow imaging experiments. A commercial 0.6 T superconductive magnet was used to test coil performance. For comparison flow imaging was also performed using a commercial set of gradient coils. Velocity profiles were obtained for several volumetric flow rates of water in tubes with inner diameters of 1.02 mm and 1.4 mm. Velocity resolution was 0.13 mm s−1, comparable to that achieved with commercial gradient coils. The quality of the velocity profile was sufficient for viscosity calculations, and thus permits future utilization of this coil package in the design of a portable MRI viscometer.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.