Abstract

A fluorescence-detection system for immunoassay applications in a laminate-based microfluidic chip has been developed and tested. Blue-LED, plastic fiber, CCD detector, mylar-D, and fluorescein are used as an excitation source, a light coupler, a fluorescence detector, a microfluidic substrate, and a sample, respectively. The system is able to detect from 1 to 1000 μg/L fluorescein. A method to avoid cross talk is also presented. © 2005 Wiley Periodicals, Inc. Microwave Opt Technol Lett 45: 456–460, 2005; Published online in Wiley InterScience (www.interscience.wiley.com). DOI 10.1002/mop.20852

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.