Abstract

We are currently developing a magnetospinography (MSG) system for noninvasive functional imaging of the spinal cord. The MSG system is a device for observing a weak magnetic field accompanied by the neural activity of the spinal cord by using an array of low-temperature superconducting quantum interference device (SQUID) magnetic flux sensors. As in the case of other biomagnetic measurement systems such as the magnetoencephalography (MEG) system, the running cost of the MSG system is mainly dependent on the liquid helium (LHe) consumption of a dewar vessel. We integrated a cryocooler into the MSG system to reduce LHe consumption. A pulse tube cryocooler with a cooling power of 0.5Wat 4 K was placed adjacent to a magnetically shielded room and was directly connected to the thermal radiation shield of the dewar by an electrically isolated transfer tube. Cold helium gas was circulated between the cryocooler and the radiation shield. Consequently, the temperature of the radiation shield decreased below 40 K. Previous studies have shown that the detection of a weak magnetic field is often hindered by severe low-frequency band noise from the cryocooler. However, the band of the MSG signals is much higher than that of the cryocooler noise. Therefore, the noise can be filtered out and has a less detrimental effect on MSG measurement than on other biomagnetic field measurements such as MEG measurement. As a result, LHe consumption was reduced by 46%, with no increase in the noise floor.