Abstract

Needle insertions have been used in several minimally invasive procedures for diagnostic and therapeutic purposes. Real-time position of the needle tip is an important information in needle steering systems. This work introduces a robot-assisted ultrasound tracking (R-AUST) system integrated with a needle shape prediction method to provide 3D position of the needle tip. The tracking system is evaluated in phantom and ex vivo beef liver tissues. An average error of 0.60 mm was found for needle insertion tests inside the phantom tissue. The R-AUST integrated with shape prediction in the beef liver tissue was able to track the needle tip with an average and maximum error of 0.37 and 0.67 mm, respectively. The average error reported in this work is within the mean allowable needle placement error (<2.7 mm) in targeted procedures. Integration of R-AUST tracking method with needle shape prediction results in a reasonably accurate real-time tracking suitable for ultrasound-guided needle insertions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.