Abstract

BackgroundHaemodynamic performance of heart valve prosthesis can be defined as its ability to fully open and completely close during the cardiac cycle, neither overloading heart work nor damaging blood particles when passing through the valve. In this perspective, global and local flow parameters, valve dynamics and blood damage safety of the prosthesis, as well as their mutual interactions, have all to be accounted for when assessing the device functionality. Even though all these issues have been and continue to be widely investigated, they are not usually studied through an integrated approach yet, i.e. by analyzing them simultaneously and highlighting their connections.ResultsAn in vitro test campaign of flow through a bileaflet mechanical heart valve (Sorin Slimline 25 mm) was performed in a suitably arranged pulsatile mock loop able to reproduce human systemic pressure and flow curves. The valve was placed in an elastic, transparent, and anatomically accurate model of healthy aorta, and tested under several pulsatile flow conditions. Global and local hydrodynamics measurements and leaflet dynamics were analysed focusing on correlations between flow characteristics and valve motion. The haemolysis index due to the valve was estimated according to a literature power law model and related to hydrodynamic conditions, and a correlation between the spatial distribution of experimental shear stress and pannus/thrombotic deposits on mechanical valves was suggested. As main and general result, this study validates the potential of the integrated strategy for performance assessment of any prosthetic valve thanks to its capability of highlighting the complex interaction between the different physical mechanisms that govern transvalvular haemodynamics.ConclusionsWe have defined an in vitro procedure for a comprehensive analysis of aortic valve prosthesis performance; the rationale for this study was the belief that a proper and overall characterization of the device should be based on the simultaneous measurement of all different quantities of interest for haemodynamic performance and the analysis of their mutual interactions.

Highlights

  • Haemodynamic performance of heart valve prosthesis can be defined as its ability to fully open and completely close during the cardiac cycle, neither overloading heart work nor damaging blood particles when passing through the valve

  • As for valve dynamics, attention has been most devoted to study the behavior in time of the valve area for both biological and mechanical prosthesis [17,18,19,20], while the leaflets motion of bileaflet mechanical heart valve (BMHV) has been somehow less investigated despite the importance of the issue [10, 21,22,23]

  • The presence of pressures crossing, in the forward flow phase, confirms the in vitro phenomena for the BMHVs known as leaflet fluttering, noticed by [30]

Read more

Summary

Introduction

Haemodynamic performance of heart valve prosthesis can be defined as its ability to fully open and completely close during the cardiac cycle, neither overloading heart work nor damaging blood particles when passing through the valve In this perspective, global and local flow parameters, valve dynamics and blood damage safety of the prosthesis, as well as their mutual interactions, have all to be accounted for when assessing the device functionality. Biological prostheses haemodynamics is usually nearly physiological but they show short durability mainly due to leaflets stiffening caused by shear stresses and calcification phenomena [6,7,8] In both cases the fluid–structure interaction plays a fundamental role in determining prosthesis functionality, a thorough analysis of flow characteristics close to the valve is essential to assess its overall performance [9]. Several literature works deal with red blood cells (RBCs) or platelets damage, providing haemolysis laws to characterize the dangerousness of the flow through the prosthetic device [35,36,37,38,39] or of the valve itself [40]

Methods
Results
Conclusion
Full Text