Abstract
Metallic magnetic microcalorimeters (MMCs) achieve energy resolution comparable to transition-edge sensors (TESs) but rely on different measurement physics that may allow MMCs to surpass TESs in some future applications. We have recently completed fabrication of new MMC γ-ray detector arrays using several exploratory sensor designs. All designs integrate the SQUID and sensor on the same chip and use a superconducting cap layer on the paramagnet, but explore different combinations of combined/separate sensing and magnetization coils and direct/flux transformer coupling to the input SQUIDs. This report describes the design and initial testing of one of these devices, which has so far demonstrated an energy resolution of 38 eV at 60 keV near 10 mK using natural-abundance silver–erbium paramagnet.
Submitted Version
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.