Abstract

This paper presents the design and fabrication of a micro Pirani gauge using VO x as the sensitive material for monitoring the pressure inside a hermetical package for micro bolometer focal plane arrays (FPAs). The designed Pirani gauge working in heat dissipating mode was intentionally fabricated using standard MEMS processing which is highly compatible with the FPAs fabrication. The functional layer of the micro Pirani gauge is a VO x thin film designed as a 100 × 200 μm pixel, suspended 2 μm above the substrate. By modeling of rarefied gas heat conduction using the Extended Fourier’s law, finite element analysis is used to investigate the sensitivity of the pressure gauge. Also the thermal interactions between the micro Pirani gauge and bolometer FPAs are verified. From the fabricated prototype, the measured device TCR is about −0.8% K−1 and the sensitivity about 1.84 × 10−3 W K−1 mbar−1.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.