Abstract
The quantitation of therapeutic antibodies by mass spectrometry often utilizes a surrogate peptide approach following enzymatic digestion of the antibody. Although this approach has been widely adopted, it is labor intensive with limited throughput in most instances. In addition, this approach can pose challenges when attempting to infer details such as quantity and modification state of the intact analyte. Recent enhancements in instrumentation and sample preparation have enabled quantitation through mass spectrometry detection of the intact protein circumnavigating many limitations of the surrogate peptide approach. Presented here is a method for quantitative analysis of therapeutic monoclonal antibodies (mAb) at the fully intact level in a complex pharmacokinetic study. This methodology yielded sensitivity down to 0.1μg/mL from 30μL of a biological sample volume to be utilized across multiple preclinical species without the need for pooling.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.