An experimental instrument for measuring a laser-induced fluorescence spectrum from a single aerosol particle is described. As a demonstration of instrument capabilities, the results of monodisperse 4.7 microm sodium chloride particles doped with fluorescent riboflavin, produced with an inkjet aerosol generator, are presented. The fluorescence of the aerosol particles is excited in the wide range from 210 to 419 nm using a pulsed, tunable optical parametric oscillator laser. The maximum of the fluorescence emission of separately measured particles is detected at 560 nm. The dependence of the fluorescence on the excitation wavelength is studied and fluorescence cross sections are estimated. Agreement between the measured fluorescence data and the literature data for riboflavin is observed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.