Abstract

The relationship between instrumental dead-time and the self-absorption coefficients, α ii, in x.r.f. matrix correction by means of influence coefficients, is not generally recognized but has important analytical consequences- Systematic errors of the order of 1% (relative) for any analyte result from experimental uncertainties in instrumental dead-time. Such errors are applied unevenly across a given range of concentration because the error depends on the calibration standards and on the instrumental conditions used. Refinement of the instrumental dead-time value and other calibration parameters to conform with influence coefficients determined elsewhere assumes exact knowledge of dead-time of the instrument used originally, and quite similar excitation conditions and spectrometer geometry for the two instruments. Though these qualifications may not be met, adjustment of any of the parameters (dead-time, reference concentration, background concentration, self-absorption and other influence coefficients) can be easily achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.