Abstract

Abstract Measurement of Pollution in the Troposphere (MOPITT) is an eight-channel gas correlation radiometer selected for the Earth Observing System AM-1 platform to be launched in 1999. Its primary objectives are the measurement of tropospheric carbon monoxide (CO) and methane (CH4). In this paper, the sensitivities of instrument signals and CO retrieval errors to various instrument parameters, especially the gas cell pressure and temperature variations, instrument radiometric noise, and ancillary data errors (such as atmospheric temperature and water vapor profile errors), are presented and discussed. In the MOPITT pressure modulator cell pressure sensitivity study, the instrument calibration process is considered, which leads to the relaxation of previous stringent requirements on the accuracy of in-orbit cell pressure monitoring. The approach of MOPITT CO retrieval error analysis is described, and the error analysis results are compared with retrieval simulation statistics. The error analysis results ...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.