Abstract

An instrument named as Volume Scattering and Attenuation Meter (VSAM) is presented. The VSAM can simultaneously measure the attenuation coefficient and the volume scattering function (VSF) from 10° to 170° with an interval of 10° at 659 nm. Using ultrapure water and NCRM-traceable polystyrene microsphere beads, the VSAM was calibrated, and the conversion factor χbθ for estimating the backscattering coefficient from the backward VSF was obtained based on Mie theory in the laboratory. For χbθ, the average relative deviation was no more than 7.77% in the range of 100°-160° between the modeled result based on VSAM and the theoretical result by Boss. Subsequently, the VSAM and ECO-VSF3 were deployed in situ in Zhanjiang Bay. The backscattering coefficient and VSF at the same angles measured by the two instruments were quite consistent. Some remarkable changes in the shape and magnitude of the VSF profile at different stations were found, with land-based pollutants composing an important suspicious source of these changes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.