Abstract

Development of the Europium-doped Strontium Iodide scintillator, SrI <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> (Eu), involves advances in crystal growth, optics and readout methodology for prototype detectors. We have demonstrated energy resolution of 3% at 662 keV for a 26 cm <sup xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sup> SrI <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> (Eu) crystal, which is equivalent to the performance obtained with Cerium-doped Lanthanum Bromide of equivalent size. Compared to standard analog readout, use of a digital readout method allows improved energy resolution to be obtained with large volume SrI <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> (Eu) crystals. Comparative gamma spectra acquired with LaBr <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">3</sub> (Ce) and NaI(Tl) quantitatively depict the value of the high resolution of SrI <sub xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</sub> (Eu) in discriminating closely spaced gamma lines for radioisotope identification applications.