Abstract
We developed an instantaneously usable screen-printed silver/silver sulfate (Ag/Ag2SO4) reference electrode with long-term stability. The reference electrode was composed of a Ag layer, a Ag/Ag2SO4 layer, an inner electrolyte layer, and a liquid junction layer printed successively on a polyimide substrate. The liquid junction and inner electrolyte layers were prepared using a silica gel–poly(vinylidene difluoride) ink. The open-circuit potentials in different electrolyte solutions were compared to that measured using a commercial reference electrode. The fabricated reference electrode stabilized the potential shift within 30 min and gave 437 mV vs. saturated KCl/Ag/AgCl measured in a 1 M KNO3 solution; the result was in good agreement with the theoretical value. Furthermore, the fabricated reference electrode remained stable for 20 days. This Ag/Ag2SO4 reference electrode showed potential for application in measurement systems where contamination by chloride ions is not desirable.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.