Abstract

The computed pressure-gradient technique has been applied to the measurement of instantaneous aortic blood flow in 15 male subjects. The mean values of those curves obtained in the ascending aorta were compared with simultaneous measurements made by the dye-dilution technique. The correlation coefficient between the two methods was 0.85 and the slope of the regression line 0.80. The ability to measure flow by the pressure-gradient technique depends mainly on the accuracy of the catheter-manometer system, radius measurement, and the estimation of “zero flow.” pressure-gradient technique; methods of blood flow measurement Submitted on June 22, 1964

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.