Abstract

Acoustic field measurements of electrohydraulic lithotripters (EHL) are typically conducted with single-element hydrophones and are subject to spatial averaging errors because the spark source location varies from shock to shock. Linear hydrophone arrays provide a means of obtaining the instantaneous sound field of EHLs and a more detailed understanding of EHL sound fields. Here, 20-element hydrophone arrays were used to study the variability of the instantaneous acoustic field of an experimental EHL. Calibrated arrays were placed at the geometric focus of an EHL and exposed to as many as 1500 shock waves using excitations of 14, 17 and 20 kV. Instantaneous data were acquired from all 20 hydrophone elements and then were processed for beamwidth, peak pressure location, and peak pressure. Instantaneous beamwidths were found to be smaller than when using a single-element hydrophone approach and peak pressures were observed to vary more as the excitation voltage increased.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.