Abstract
A new method is developed for instantaneous α-determination in the 1/E1+α epithermal neutron spectrum. It is based on coirradiation of three bare resonance detectors, followed by gamma-counting on a calibrated Ge(Li)-detector. The technique is applied in two channels of the Thetis reactor (Gent) and the results are critically compared with those obtained by other methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.