Abstract

The integration of a shape memory actuator is a potential mechanism to achieve a consistent perimodiolar position after electrode insertion during cochlear implant surgery. After warming up, and therefore activation of the shape memory effect, the electrode array will change from a straight configuration into a spiral shaped one leading to a final position close to the modiolus. The aim of this study was to investigate whether the integration of an additional thin wire (referred to as an "inlay") made of Nitinol, a well-established shape memory alloy, in a conventional hearing preservation electrode array will affect the insertion behaviour in terms of increased risk of insertion trauma. Six conventional Hybrid-L electrode arrays (Cochlear Ltd., Sydney, Australia) were modified to incorporate a wire inlay made of Nitinol. The diameter of the wires was 100µm with a tapered tip region. Electrodes were inserted into human temporal bone specimens using a standard surgical approach. After insertion and embedding in epoxy resin, histological sections were prepared to evaluate insertion trauma. Insertion was straightforward and no difficulties were observed. The addition of a shape memory wire, thin but also strong enough to curl the electrode array, does not result in histologically detectable insertion trauma. Atraumatic insertion seems possible.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.