Abstract

The purpose of this study was to evaluate the feasibility of our innovative, replaceable heart valves that can be easily detached from the sewing ring at the time of repeat replacement. The prototype devices consist of the base magnet ring assembly and the valve magnet ring assembly that utilize magnetic coupling force for the locking mechanism. Magnetic coupling strength was evaluated in vitro. Prototype bioprosthetic valves were implanted acutely in three sheep to confirm the feasibility of the replaceable mitral valve. The static separation force of prototype size #25 was 12.5 lb, meeting the design goal. In situ attachment and detachment of the valve magnet ring assembly from the base magnet ring assembly were very easily accomplished in all animals. The magnetic coupling did not decouple even under extremely high left ventricular pressures. We have demonstrated the feasibility of this innovative concept of a replaceable mitral valve.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.