Abstract

A novel oscillator circuit with automatic capacitance compensation (ACC) capability has been coupled with QCM sensors coated with quinoxaline and pyrazine bridged cavitands to detected aromatic and chlorinated compounds in water. With double-side immersed 10-MHz crystals coated with the quinoxaline cavitand, the detection of toluene in deionized water was possible down to concentrations of few parts per million. The ACC oscillator advantageously provides the simultaneous measurement of the sensor resonant frequency, damping, and value of the compensated parallel capacitance. This enabled to observe that the analyte sorption in the cavitand coating not only brings about a mass uptake, but also an increase of losses, and, apparently, a rise in the coating average permittivity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.