Abstract

Background Intraoperative blood loss has been shown to be an important factor correlating with morbidity and mortality in liver surgery. A 5-cm long instrument with variably deployable metal electrodes using in-line radiofrequency ablation (ILRFA) energy was used for hepatic transection in an attempt to reduce bleeding. Methods Eight patients underwent liver resection. At each resection, half the resection was performed with ILRFA and the other half was performed with an ultrasonic aspirator alone. Blood loss was measured for each mode of resection. Results The mean blood loss using ILRFA was 6.5 (±3.7) mL/cm 2 compared with 20.4 (±8.7) mL/cm 2 by using the ultrasonic aspirator ( P = .004). Conclusions In-line radiofrequency ablation reduced bleeding during hepatic parenchymal transection when compared with the ultrasonic aspirator.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.