Abstract

This study reports a simple, rapid, low-cost, robust, and multiplexed barcoded paper-based assay (BPA) compatible with mobile devices. An inkjet printer and an XYZ dispensing platform were used to realize mass-manufacturing of barcoded paper-based analytical devices (BPADs) with high precision and efficiency. We designed a new group of barcodes and developed an application (APP) for the reading of the new code. The new barcodes possess a 16 times higher coding capacity than the standard Codabar code in our experiment on drug residue detection. The BPA system allows applications in the assays of blood-transmitted infections, drug residues in milk and multiplex nucleic acids. The whole detection process and the readout of the results can be completed within 10 minutes. The limit of detection for enrofloxacin (ENR) (8 ng mL-1) satisfies the requirements of drug residue monitoring. Its high rapidity, simplicity, efficiency and selectivity make the BPA system extremely suitable to be applied in rapid and on-site detection.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.