Abstract

This article reports, for the first time, the results of detailed research on the application of inkjet 3D printing for the fabrication of microfluidic structures. CAD designed test structures were printed with four different printers. Dimensional fidelity, shape conformity, and surface roughness were studied for each printout. It was found that the minimum dimension (width or depth) for a properly printed microfluidic channel was approximately 200 μm. Although the nominal resolution of the printers was one order of magnitude better, smaller structures were significantly deformed or not printed at all. It was also found that a crucial step in one-step fabrication of embedded microchannels is the removal of the support material. We also discuss the source of print error and present a way to evaluate other printers.The printouts obtained from the four different printers were compared, and the optimal printing technique and printer were used to fabricate a microfluidic structure for the spectrophotometric characterisation of beverages. UV/VIS absorbance characteristics were collected using this microfluidic structure, demonstrating that the fabricated spectrophotometric chip operated properly. Thus, a proof-of-concept for using inkjet 3D printing for the fabrication of microfluidic structures was obtained.