Abstract
A fast automated system for rapid electrophoretic separations in short conventional capillaries employing contactless conductivity detection is presented. The instrument is based on pneumatic pressurization and does not require a conventional pump. The required pressures and flow rates for the different steps of the injection and flushing processes are produced with the help of two flow restrictors. The device is implemented on a microfluidic breadboard with dimensions of ca. 13 × 20 cm and employs miniature valves. Nine inorganic cations, namely NH4+, K+ , Na+ , Ca2+ , Mg2+ , Mn2+ , Sr2+ , Li+ , and Ba2+ , could be separated in a capillary of 10 μm inner diameter and 6 cm effective length within 25 s. Following a reduction of the effective length to 4cm, still five inorganic cations could be separated in a time span of 12s. The repeatability of peak areas was better than 3.1% and limits of detection between 3.5 and 5.5 μM were achieved.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.