Abstract

The initial axial velocities for the positive and negative cytochrome c molecule-ions desorbed from a liquid matrix (3-nitro-benzyl alcohol) using a 17 ns ultraviolet laser (266 nm) have been determined. The method employed was based upon measurement of ion flight times through a field-free path co-axial with the ion optical axis. The possible interferences, such as energy deficits, due to the presence of electric fields were avoided by using a grid-electrode placed at a very short distance from the sample stage. The potential of this electrode was maintained at the same potential as the sample stage, thereby creating a narrow field-free region within which ion generation took place. Systematic variations of the potential at a second grid-electrode caused a gradual shift of the ion flight times, which could be related to the initial ion velocities. The ion velocities were obtained by correlating the measured time-shifts with values from theoretical analysis. The positive and negative molecule-ions were found to have axial velocities of 840±70 and 750±40 ms −1, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.