Abstract
Clinical samples were evaluated with the Mobile Analysis Platform (MAP) to determine platform performance for detecting respiratory viruses in samples previously characterized using clinical reverse transcriptase polymerase chain reaction assays. The percent agreement between MAP and clinical results was 97% for influenza A (73/75), 100% (21/21) for influenza B, 100% (6/6) for respiratory syncytial virus (RSV), and 80% (4/5) for negative specimens. The approximate limit of detection of the MAP was 30 copies/assay for RSV and 1500 copies/assay for Middle East respiratory syndrome coronavirus.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.