Abstract
Endovascular repair has become the standard of care for treatment of abdominal aortic aneurysms. The endografts and delivery systems for endovascular aneurysm repair have undergone multiple generations of technologic advancements. However, a significant remaining challenge for a satisfactory long-term outcome is to improve the performance of these devices in nonideal proximal sealing zones. In particular, short (<15 mm) and highly angulated (>60 degrees) necks can threaten long-term exclusion of the aneurysm even with the current generation of endografts. One of the main reasons for proximal infrarenal neck failure is the inability to accurately position the endograft precisely below both renal arteries. This is a report of the first-in-human implantations of the GORE EXCLUDER Conformable AAA Endoprosthesis (W. L. Gore & Associates, Flagstaff, Ariz), an investigational device, in anatomies with standard neck lengths and angulation. This device has been designed to provide repositionability, conformability, and, for the first time, optional angulation control. This device is commercially available in Europe.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Vascular Surgery Cases and Innovative Techniques
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.