Abstract
This study evaluated the resistance to gapping and the mode of failure for 2 knotless suture anchor systems used for rotator cuff repair compared with the performance of a conventional titanium anchor system. Eight matched pairs of fresh-frozen humeri were dissected free of all soft tissues and scanned to measure bone mineral density (BMD). The suture anchor systems tested were the TwinFix 5.0 Titanium (Smith & Nephew, Andover, MA), Bioknotless RC (DePuy Mitek, Norwood, MA), and Magnum (Opus Medical, San Juan Capistrano, CA), and each was inserted into each humerus. Cyclic, tensile loading was applied through the suture loop for 5000 cycles, or until failure, by using a servohydraulic testing machine. Gapping distances, defined as increasing elongation of the bone/anchor/suture system, were continuously measured. Total cycles to failure and mechanism of failure were documented. Mean initial (first cycle) and final (last cycle) gapping distances were 3.81 mm and 5.36 mm for the TwinFix 5.0, 4.02 mm and 5.34 mm for the Bioknotless RC, and 3.56 mm and 4.98 mm for the Magnum anchors. No significant difference was detected among mean gap openings (P > .05). However, the Bioknotless RC had more early failures (5) than the other 2 implants (1 each), approaching significance (P = .07). Trials of the Bioknotless RC that did not fail early were found to have significantly less gap opening than the other 2 systems for both initial (1.89 mm vs 3.82 mm for the TwinFix 5.0 and 3.56 mm for the Magnum) and final (2.00 mm vs 4.68 mm for the TwinFix 5.0 and 4.24 mm for the Magnum) gap opening. BMD was a significant predictor of initial (P = .029) and final (P = .008) gap opening, whereas the site of anchor insertion was a significant predictor of final displacement. The Opus Magnum was comparable with a conventional suture anchor, but the Mitek Bioknotless RC showed a trend toward early failure. Biomechanical analysis of knotless suture anchor systems can demonstrate trends among implants in an experimental setting. Knowledge of these trends could influence implant selection.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.