Abstract
We describe our initial experience with the placement of two premounted balloon expandable intracranial Jostent stent-grafts within the intracavernous internal carotid artery for the treatment of a symptomatic large intracavernous aneurysm in one case and a post-traumatic caroticocavernous fistula in the second. Among the initial technical complications we encountered were stent-graft migration and rapidly progressive intragraft thrombosis, with delayed sealing of the stent-graft coverings and exclusion of the pathologies relating to the use of abciximab in both cases. Despite these initial problems both cases had excellent short-term clinical outcomes with angiographic exclusion of both lesions by day three and good clinical and angiographic outcomes at one and two months respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.