Abstract

Fast low-angle multiecho (FLAME) imaging uses partial flip angles of less than 90 degrees with 180 degrees radiofrequency refocusing pulses. The partial flip angle permits imaging with shorter repetition time (TR) values on the order of 750-1,000 ms for 30 degrees angles with image contrast characteristics identical to those obtained with conventional 90-180 degrees schemes and TRs on the order of 2,500 ms. The approximately threefold reduction in imaging time is accompanied by a decrease in signal-to-noise ratio. In many circumstances, however, this trade-off may produce entirely acceptable images of the CNS at a significant reduction in imaging time.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.