Abstract

There has been reluctance to use intraoperative transesophageal echocardiography (TEE) in small infants. We assessed the utility and safety of a new miniaturized multiplane micro-TEE probe in small infants undergoing cardiac operations. Hemodynamic and ventilation variables were prospectively recorded before and after micro-TEE insertion and removal in infants weighing 5 kg or less undergoing cardiac operations. The study included 42 patients with a mean weight of 3.6 +/- 0.9 kg (range, 1.7 to 5 kg). All probe insertions were successful. There were no complications or clinically significant changes in hemodynamic or ventilation variables. Information provided by TEE resulted in surgical revision in 6 of the 42 patients. The micro-TEE provides high quality, useful diagnostic images without hemodynamic or ventilation compromise in small infants undergoing cardiac operations. This advance is important with the growing trend towards complete repair of complex structural heart disease in small infants.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.