Abstract

We evaluated the effectiveness of a novel bicompartmental knee arthroplasty prosthesis for the treatment of degenerative disease affecting the medial and patellofemoral compartments. The study included 36 knees in 32 patients with a mean follow-up of 21 months. The mean Knee Society functional survey and Western Ontario McMaster Osteoarthritic Index Survey scores were 65.4 and 75.8, respectively. Thirty-one percent of patients were unsatisfied with the surgery, and 53% stated that they would not repeat the surgery. We report an overall survival rate of 86% with one catastrophically failed tibial baseplate. We conclude that this prosthesis provides inconsistent pain relief and unacceptable functional results for bicompartmental arthritis. Short-term survival of this prosthesis was unacceptably low, and therefore, we no longer implant it at our institution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.