Abstract

To investigate the dose calculation accuracy in water medium of the Advanced Collapsed cone Engine (ACE) for three sizes of COMS eye plaques loaded with low-energy I-125 seeds. A model of the Oncura 6711 I-125 seed was created for use with ACE in Oncentra® Brachy (OcB) using primary-scatter separated (PSS) point dose kernel and Task Group (TG) 43 datasets. COMS eye plaque models of diameters 12, 16, and 20mm were introduced into the OcB applicator library based on 3D CAD drawings of the plaques and Silastic inserts. To perform TG-186 level 1 commissioning, treatment plans were created in OcB for a single source in water and for each COMS plaque in water for two scenarios: with only one centrally loaded seed, or with all seed positions loaded. ACE dose calculations were performed in high accuracy mode with a 0.5×0.5×0.5mm3 calculation grid. The resulting dose data were evaluated against Monte Carlo (MC) calculated doses obtained with MCNP6, using both local and global percent differences. ACE doses around the source for the single seed in water agreed with MC doses on average within <5% inside a 6×6×6cm3 region, and within <1.5% inside a 2×2×2cm3 region. The PSS data were generated at a higher resolution within 2cm from the source, resulting in this improved agreement closer to the source due to fewer approximations in the ACE dose calculation. Average differences in both investigated plaque loading patterns in front of the plaques and on the plaque central axes were ≤2.5%, though larger differences (up to 12%) were found near the plaque lip. Overall, good agreement was found between ACE and MC dose calculations for a single I-125 seed and in front of the COMS plaques in water. More complex scenarios need to be investigated to determine how well ACE handles heterogeneous patient materials.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call