Abstract

An accelerator-based boron neutron capture therapy (AB-BNCT) system was installed at the Shonan Kamakura General Hospital (SKGH). We confirmed that a stable operation was possible for 1 h at a current of 30 mA. The evaluated thermal neutron flux was 2.8 × 109 cm−2 s−1 and in good agreement (±5%) with the calculated values. The daily variation was within ±2%. The ambient dose rate due to residual radioactivity after irradiation was approximately 5 μSv/h using a lead shutter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.