Abstract

The Palmaz-Schatz (PS) stent has effectively reduced restenosis; however its rigidity makes it sometimes difficult or impossible to deliver. The initial and follow-up outcomes with the ACS Multi-Link stent (deployed from April to November 1995) were evaluated in 70 patients (79 lesions): unplanned in 34% (abrupt closure 1%; threatened closure 5%; suboptimal results 28%) and planned in 66%. Three to six month follow-up angiograms were analyzable in 67 lesions; 96% procedural (in nine lesions PS stenting had failed) and 95% clinical success were achieved. In-hospital mortality was 1.4%. Myocardial infarction occurred in 2.9%, and subacute stent thrombosis in 1.4%. Stenting improved immediately the minimal luminal diameter (from 0.97+/-0.41 mm to 2.72+/-0.31 mm), but at 6 months it had decreased to 1.89+/-0.44 mm. Angiographic restenosis (<50% diameter stenosis) occurred in 11, a rate of 16.4%; target lesion revascularization (TLR) was required in six (re-PTCA in five or bypass surgery in one; 6/67=8.7%). Actuarial 1-2 year survival rate was 91%, 80% surviving free from major complications or need for TLR. We conclude that the ACS Multi-Link stent can be implanted in lesions unsuited for the PS stent with a high success rate, and an anticipated restenosis rate perhaps comparable to with the PS stent.