Abstract

Light pipe infrared spectra of adsorbed layers and powders are shown. They are compared with specular external reflection, ATR, KBr pellet transmission, and diffuse reflection spectra. The resultant light pipe spectra are found to obey the predictions of the theory presented in the previous companion paper [Appl. Spectrosc. 41, 46–62 (1987)], including the high sensitivity and S/N predicted to be attainable. They are easy to obtain and to interpret. Only fractions of a milligram of sample are required in favorable cases. The light pipe is convenient as a reaction cell, allowing changing spectra to be followed in real time. In the previous paper were presented equations for absorption spectra of thin layers of material on the inner surface of a light pipe, in the infrared. In the present paper the equations are tested against experimental reality. These tests are a crucial verification of the theory.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.