Abstract

A method for monitoring hemodialysis based on quantitative infrared spectroscopic determination of the molecules dialyzed from patient blood is reported. The measurements are reagent-free and aim at real-time and in-line monitoring of the hemodialysis patient. A flow cell using attenuated total reflection infrared spectroscopy is coupled downstream of the dialysis filter unit. A calibration model has been developed from real hemodialysis samples analyzed by chemical reference analysis and from artificially mixed dialysis samples. The infrared monitoring of hemodialysis includes quantitative determination of urea as the lead substance, as well as glucose, lactate, and creatinine, all at a precision only limited by the chemical reference analysis. The flow cell can be fitted to all standard hemodialysis systems. Preliminary tests with hemodialysis patients have demonstrated that detoxification can be clearly monitored. Furthermore, these experiments demonstrate that a wide, real-time control of the patient's physiological parameters is possible with this method, which could lead to increased patient safety.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.