Abstract

Combination of ultraviolet (UV) photolysis with infrared (IR) spectroscopy (or UV/IR for abbreviation) is a powerful tool to study various chemical photoreactions, while cryostat and sample-cell windows define the working ranges for both UV and IR beams. Although diamond window has a very wide transmission range from UV to IR, the extreme cost, the absorptions at 1800–2600cm−1 and other problems prevent it from being the solution for all cases. In this paper, a gas-exchange cryostat was modified to realize a UV/mid-IR experiment at cryogenic temperatures. Several windows (including diamond) were discussed as options. A di-nitrogen iron complex trans-[Fe(DMeOPrPE)2(N2)H][BPh4] [DMeOPrPE=1,2-bis(dimethoxypropylphosphino)ethane] was studied as a real photolysis example. Alternatively, a cold-finger cryostat was modified for UV/far-IR compatible experiments. Non-photolysis samples K5[Mo4O11(R,S-Hhomocit)2]Cl·5H2O (H4homocit=homocitric acid) and [(n-Bu)4N]2[Fe4S4(PPh)4] were studied at cryogenic temperatures. Sample cell windows can also be used as a natural way for choosing photolysis wavelength (in addition to the use of optical filters).

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call