Abstract

Infrared (IR) Focal plane array (IRFPA) cameras are nowadays both, more accessible and with a broad variety in terms of detectors design. In many cases, the IRFPA characterization is not completely given by the manufacturer. In this paper a long wave 8-12 [μm] microbolometer IRFPA is characterized by means of calculating the Noise Equivalent Temperature Difference (NETD) and the Correctability performance parameters. The Correctability parameter has been evaluated by using a black body radiator and Two-Points calibration technique. Also, the Transfer Function of the microbolometer IR camera has been experimentally obtained as well as the NETD by the evaluation of radiometric data from a blackbody radiator. The obtained parameters are the key for any successful application of IR imaging pattern recognition.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.