Abstract

Cryogen spray cooling (CSC) is an effective method to minimize epidermal damage during laser treatment of various cutaneous anomalies such as port wine stains, excess hair, and facial rhytides. Radiometric temperature measurements provide a noninvasive method to estimate the skin surface temperature. Since the infrared absorption spectrum of the cryogen film has remained unknown, assumptions for those values may lead to inaccurate temperature estimations. We have constructed several high-pressure infrared transparent cuvettes to determine the absorption coefficient of room temperature R-134a in liquid phase using Fourier Transform Infrared Spectroscopy (FTIR) in the 2.5 - 14 μm spectral bandwidth. Results demonstrate that liquid R-134a has several absorption bands in the infrared, with those between 7 - 10.5 and 11.5 - 12.5 μm being the most prominent. Additionally, the absorption coefficient at two common radiometric bands, 3 - 5 and 7 - 11 μm differ by four orders of magnitude. Results of this study will lead to further improvements in interpreting radiometric temperature measurements when using CSC.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.