Abstract

Our recent diffuse optical tomography experiments on human lower legs and forearms are presented using the time-resolved measuring system and image reconstruction algorithm based on the modified generalized pulse spectrum technique. It was shown that the spatial resolution and quantitativeness of the resultant images, was rather poor, and the interior blood vessels invisible in the absorption images. To clarify this issue, the influences of target contrast and size on the image reconstruction were investigated with simulated data. We have found that the quantitativeness of the reconstructed optical properties was prone to be spoiled by the small size ratio and high contrast of the interior targets to the background, and the incompleteness of information embedded in the featured data-types, evidently answers for the degradation of the image quality. It was shown in a further simulative investigation that the image quality could be substantially improved by making full use of the time-resolved data.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.