Abstract

Although useful for visualizing blood flow during revascularization surgery, the permeability of near-infrared fluorescence (NIR) angiography using indocyanine green (ICG) does not allow for vessel stenosis visualization. We hypothesized that changes in ICG fluorescence intensity reflect vessel stenosis, and evaluated the influence of stenosis on blood flow by ex vivo experimentation. The vessel stenosis model comprised a silicon tube, a graft occluder, and artificial blood. During near-infrared angiography, the fluorescense intensity was calculated during pre- and post-stenosis of an artificial circuit, using a NIR angiography. We measured the maximum fluorescence intensity and the time to maximum fluorescence intensity. Severe stenosis (≥75%) attenuated the increase in ICG fluorescence intensity in the tube significantly, pre- and post-stenosis. The time to maximum fluorescence intensity did not differ between sites pre- and post-stenosis, irrespective of stenosis severity. Stenosis affected the ICG fluorescence intensity through the vessel. Thus, quantitative analysis using NIR angiography may detect severe vessel stenosis (≥75%), and the extinction curve of indocyanine fluorescence intensity may support the evaluation of blood flow. The absence of differences in the time to maximum fluorescence intensity for degrees of stenosis might suggest a limitation of previous conventional qualitative assessments.