Abstract
A range of preliminary designs of customised total knee implants (CTKIs) was created by resurfacing the distal femur and applying different tibial bearing surface curvatures. These were then compared with a scaled off-the-shelf symmetric total knee implant (STKI). To evaluate the biomechanical performance, a dynamic knee simulation model was created with patient-specific muscle and ankle joint loads calculated from an OpenSim musculoskeletal model. Simulation results showed the transverse curvatures of the tibial bearing surface influenced femoral mediolateral translation, while its longitudinal curvatures affected femoral adduction. Compared to the STKI, the CTKIs could restore patient knee function.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Computer Methods in Biomechanics and Biomedical Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.