Abstract

Commercial transparent glass has received attention in relation to its use as a gamma radiation detector. This kind of material presents a strong post-irradiation thermal decay at room temperature, which is a problem for high dose dosimetry purposes. This effect may be avoided by taking measurements always at the same time after irradiation of the glass samples. However, in this work effort has been given to this material characteristic, submitting the glass samples to different thermal treatments. The best results were obtained with a post-irradiation treatment at 130 degrees C during 10 min. This procedure seems to destroy the unstable entities responsible for the initial strong decay of the sample response, showing acceptable stability as a function of room temperature storage time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.