Abstract

The tearing of the fibers of biomaterials employed in implants or bioprostheses leads to early the failure of these devices. The purpose of this study was to determine the force necessary to propagate a tear in a biological tissue, calf pericardium, when sutured. We analyzed the outcome of 230 trials. There was a loss of resistance to tearing in samples sutured edge-to-edge as compared to unsutured control samples. This loss was not observed when the suture was preceded by an intact or protective zone. The values corresponding to the tearing force for an overlapping suture, especially when sewn with Gore-Tex(R), were higher than those obtained in controls. This study confirms the deleterious effect of the edge-to-edge suture, which can be minimized by protecting the suture, and the excellent behavior of the overlapping suture.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.